Hamilton Medical Recall

Buckeye is in receipt of the below notice from ODM.  We understand many of you have already seen and your Clinical Engineering departments have acted on this notice, but Buckeye wants to ensure you are aware:

ODM would like to make the MCEs aware of the below critical recall from the FDA.   

The full announcement is available on the FDA's website

Hamilton Medical Inc. Recalls HAMILTON-C1, T1, MR-1 Ventilators for Capacitator Leaks and Short Circuits

Hamilton Medical, Inc. is recalling the HAMILTON-C1, T1, MR-1 ventilators because the capacitators may leak electrolyte fluid onto the ventilator’s control board. If the control board contacts the electrolyte fluid, the control board or installed spare parts could short circuit. As a result of the short circuit, the ventilator may switch to “Ambient State.”

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Questions?

If you have questions about this recall, contact Hamilton Medical Inc. at 1-800-426-6331 or email reno.techsupport@hamiltonmedical.com.